The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.


Metrics

 
NVD enrichment efforts reference publicly available information to associate vector strings. CVSS information contributed by other sources is also displayed.
CVSS 4.0 Severity and Vector Strings:

NVD assessment not yet provided.

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to [email protected].

URL Source(s) Tag(s)
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 CVE, ICS-CERT Third Party Advisory  US Government Resource 

Weakness Enumeration

CWE-ID CWE Name Source
CWE-798 Use of Hard-coded Credentials πŸ‘ cwe source acceptance level
NIST  
CWE-259 Use of Hard-coded Password ICS-CERT  

Known Affected Software Configurations Switch to CPE 2.2

CPEs loading, please wait.

Denotes Vulnerable Software
Are we missing a CPE here? Please let us know.

Change History

4 change records found show changes

CVE Modified by ICS-CERT 6/16/2026 10:51:20 PM

Action Type Old Value New Value
Added Affected
[{"vendor":"n/a","product":"Baxter Sigma Spectrum Infusion Pumps","versions":[{"version":"Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24","status":"affected"}]}]


CVE Modified by CVE 11/20/2024 11:59:10 PM

Action Type Old Value New Value
Added Reference
https://www.us-cert.gov/ics/advisories/icsma-20-170-04


CVE Modified by ICS-CERT 5/14/2024 2:19:57 AM

Action Type Old Value New Value

Initial Analysis by NIST 7/08/2020 11:08:32 AM

Action Type Old Value New Value
Added CVSS V3.1
NIST AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H


Added CVSS V2
NIST (AV:N/AC:L/Au:N/C:P/I:P/A:P)


Added CWE
NIST CWE-798


Added CPE Configuration
AND
 OR
 *cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*
 OR
 cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*
 cpe:2.3:h:baxter:wireless_battery_module:17:*:*:*:*:*:*:*
 cpe:2.3:h:baxter:wireless_battery_module:20d29:*:*:*:*:*:*:*
 cpe:2.3:h:baxter:wireless_battery_module:20d30:*:*:*:*:*:*:*
 cpe:2.3:h:baxter:wireless_battery_module:20d31:*:*:*:*:*:*:*
 cpe:2.3:h:baxter:wireless_battery_module:22d24:*:*:*:*:*:*:*


Changed Reference Type
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 No Types Assigned


https://www.us-cert.gov/ics/advisories/icsma-20-170-04 Third Party Advisory, US Government Resource